6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1322

■■■■■■■■

tiamo 2.2 (ProcessLab)

for ISE (+3)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

23.0 mV

for ISE (-3)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-16.0 mV

for ISE (+4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

17.0 mV

for ISE (-4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-12.0 mV

pH(0) monitoring (for pH electrodes only)
on | off (Default value: off)

If this option is enabled then a check will be made to see whether the elec-
trode zero point pH(0) determined during the calibration of a pH electrode
is within the defined limits. If it is outside them, a corresponding message
appears and you can select whether you will nevertheless accept the value
or whether the previous value is to be retained.

Lower limit

Lower limit for pH(0).

Range

-2.147E+9 ... 2.147E+9

Default value

6,750

Upper limit

Upper limit for pH(0).

Range

-2.147E+9 ... 2.147E+9

Default value

7,250

E(0) monitoring (for ISE electrodes only)
on | off (Default value: off)

If this option is enabled then a check will be made to see whether the elec-
trode zero point E(0) determined during the calibration of an ISE electrode
is within the defined limits. If it is outside them, a corresponding message
appears and you can select whether you will nevertheless accept the value
or whether the previous value is to be retained.