Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1082

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1066

■■■■■■■■

tiamo 2.2 (ProcessLab)

730, 778, 789
Range

0.001 ... 999.999 mL

Default value

10.0000 mL

774
Range

0.01 ... 999.999 mL

Default value

10.0000 mL

Dosing rate

The rate at which the volume is dosed. The maximum dosing rate depends
on the cylinder volume of the exchange unit or dosing device used. If the
entered dosing rate is too high for the selected dosing device, it will auto-
matically be reduced during dosing to the largest possible value.

Titrando, 846, 814, 815, 855, 864, 874
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

730, 774, 778, 789
Range

0.01 ... 160.00 mL/min

Default value

maximum mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

maximum mL/min

Note

The dosing rate should be reduced for viscous liquids.

Filling rate

The buret will be filled at this speed after the dosing. The maximum filling
rate depends on the cylinder volume of the exchange unit or dosing device
used. If the entered filling rate is too high for the selected dosing device, it
will automatically be reduced during dosing to the largest possible value.

Titrando, 814, 815, 855, 864, 874
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

730, 774, 778, 789
Range

0.01 ... 160.00 mL/min

Default value

maximum mL/min