5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1097

off
Stops when the preset time has elapsed since the start of the dosing.

Filling rate

Rate at which the dosing cylinder is to be filled after dosing. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
device used. If the entered filling rate is too high for the selected dosing
device, it will automatically be reduced during dosing to the largest possible
value.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

maximum mL/min

Note

The filling rate should be reduced for viscous liquids.

5.6.6.3.5.5

DOS U - Monitoring

Tab: Methods

DOS U Properties... Monitoring

Command name

Name of the command.

Input

25 characters

The following monitoring processes can be switched on and defined on this
tab:

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is switched on then the measured value will be monitored and
any exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is triggered.