Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 783

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

767

Max. rate

Rate at which dosing is carried out outside of the Control range. The max-
imum dosing rate depends on the cylinder volume of the exchange unit or
dosing unit used.

Titrando
Range

0.01 ... 166.00 mL/min

Default value

1.00 (slow), maximum (optimal), maximum
(fast) mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

1.00 (slow), maximum (optimal), maximum
(fast) mL/min

Min. volume increment

This parameter determines the volume increment which is added at the start
and finish of the titration. The smaller the increment that is selected, the
slower the titration and the higher the accuracy.

Range

0.1 ... 9.9 µL

Default value

minimum (slow), minimum (optimal), 5.0
(fast) µL

Stop criterion

Stop criterion

The titration us stopped after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be stopped
when a particular drift has been achieved or after a preset time.

Titrino
Selection

drift | time

Default value

drift

Titrando
Selection

drift | rel. drift | off | time

Default value

drift

drift
The titration is stopped when the stop drift has been reached.
rel. drift
The titration is stopped when the relative stop drift (drift value at time of
start + relative stop drift) has been reached.
off
The titration will not be stopped.