Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1020

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1004

■■■■■■■■

tiamo 2.2 (ProcessLab)

Solution

Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Note

If an existing solution is selected as the solution to be added then the
Concentration and Unit of this solution will be adopted for calculating
the standard addition. If not defined is selected, then Concentration
and Unit must be entered on the Standard addition tab.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando, 867
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Sensor

Selection of a sensor of type ISE electrode from the sensors configured in
the sensor table.

Selection

'Sensor name' | ISE electrode

Default value

ISE electrode

Temperature measurement

Type of temperature measurement.