5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

860

■■■■■■■■

tiamo 2.2 (ProcessLab)

Fix endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value
, time or volume) for the fix endpoint from the measuring point list.

[New]

Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.8.7.10, page 862).

[Properties]

Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.8.7.10, page 862).

[Delete]

Deletes the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and temperature
for the minimum measured value will be interpolated from the measuring
point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

1.0 ... 2000.0 mV/mL

Default value

25.0 mV/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and temperature
for the maximum measured value will be interpolated from the measuring
point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.