Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1315

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1299

Maximum Dosing/Filling rate for exchange unit as a function of the
cylinder volume:

Cylinder volume

Maximum rate

1 mL

3.0 mL/min

5 mL

15.0 mL/min

10 mL

30.0 mL/min

20 mL

60.0 mL/min

50 mL

150.0 mL/min

Note

Enter lower rates for high-viscosity liquids.

Tubing parameters
Defines the length and diameter of the tubing connected to the exchange
unit.

Note

Default values have already been entered for the tubing parameters; these
correspond to the dimensions of the standard tubing supplied. If you do
not make any alterations to the tubing connections then you do not need
to modify the tubing parameters. For information on length and diameter
of other tubing see Products/Accessories and columns under

http://

products.metrohm.com

.

Dosing tip

Length

Length of the tubing connecting the cock to the dosing tip.

Range

0.0 ... 999.9 cm

Default value

40.0 cm

Diameter

Diameter of the tubing connecting the cock to the dosing tip.

Range

0.0 ... 9.9 mm

Default value

2.0 mm