5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

692

■■■■■■■■

tiamo 2.2 (ProcessLab)

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.4.4.10

SET U - Fix endpoint evaluation

Dialog window: Method

SET U Properties... Additional evaluations Fix end-

point evaluation

[New]/[Properties] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.4.5

SET Ipol

5.6.3.4.5.1

SET Ipol - Overview

Dialog window: Method

SET Ipol Properties... SET Ipol - 'Command name'

Command for Set Endpoint Titrations with voltametric measurement
(selectable polarization current).

Devices
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 841, 842, 852, 857, 888, 901, 902, 904,
905, 906, 907

Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799

Robotic Titrosampler: 855