Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 143

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Workplace

tiamo 2.2 (ProcessLab)

■■■■■■■■

127

Note

The following sample data fields can no longer be edited if they were
defined as Fixed value in the method (see Chapter 5.6.2.2.5, page
427)
.

Sample position

Position of the sample on the rack. This number is used to move to the
sample position with the command MOVE. For this to be achieved, the
Sample position option must be set under Target for the Move para-
meter.

Range

1 ... 999 (Increment: 1)

Default value

'empty'

ID1…ID16

Sample identifications ID1…ID16. Any text may be entered into this field,
the checking for type and limit values does not take place until the start of
the determination. For sample identifications of the type Date/Time, the
entry must be made in the format YYYY-MM-DD or YYYY-MM-DD
hh:mm:ss
. If text templates are defined for the sample identification, then
these can be selected (see Chapter 3.5.1, page 120).

Input

100 characters

Selection

Selection of text templates

Sample size

Sample size (initial weight). Only numbers, decimal points, +, –, E and e can
be entered in this field. The checking for limit values does not take place
until the start of determination.

Note

If a value is entered or imported into this field, then in addition the time
of the data entry and the data source are stored manually, while the
'Name of the scale', the 'Name of the barcode reader' or the
'Name of the import file') are stored automatically at the same time,
placed in the determination and identified in the Information sub-
window as Data entry and Data source.