6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1298

■■■■■■■■

tiamo 2.2 (ProcessLab)

IEU
Input

8 characters

Default value

'empty'

EU
Input

10 characters

Default value

'empty'

Parameters for preparation
Configures the parameters to be used in the command PREP.

Volume

Volume to be dosed in during preparation.

Range

0.00000 ... 99999.99999 mL

Selection

Cylinder volume

Default value

Cylinder volume

Cylinder volume
The whole contents of the cylinder will be dosed in.

Cycles

Number of rinsing cycles during preparation.

Range

1 ... 9

Default value

2

Dosing rate

Speed at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out the speed will automatically be reduced to
the largest possible value.

Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Filling rate

Speed at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the
function is carried out the speed will automatically be reduced to the largest
possible value.

Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum