5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1118

■■■■■■■■

tiamo 2.2 (ProcessLab)

Selection

Device types | 814 USB Sample Processor

Default value

814 USB Sample Processor

Target

Tower

Selection of the tower of the Sample Processor for moving to the required
lift position. Both Tower 1 and Tower 2 can always be selected even if the
Sample Processor has only one tower.

Selection

1 | 2

Default value

1

Move

Selecting the target position on the rack:

Selection

Sample position | Rack position | Special
beaker
| Relative angle

Default value

Sample position

Sample position
Position defined for the current determination in the Run subwindow of
the program part Workplace as the Sample position parameter.
Rack position
Absolute position on the rack.
Special beaker
Special position on the rack defined in the rack properties of the Con-
figuration
program part.
Relative angle
The sample rack can be rotated by a certain angle relative to the current
position independently of rack positions. This can be used for example
to remove vial caps automatically.

Number

Specification of the number of the rack position. This parameter is only visible
for Move = Rack position or Special beaker. The rack position can be
entered as a formula as well, which is necessary for the automatic calibration
of electrodes, for example. The formula editor is opened by right-clicking on
the input field.

Move = Rack position
Range

1 ... 999

Default value

1

Move = Special beaker (for 778, 789, 814, 815, 855, 864, 874)
Range

1 ... 16

Default value

1