4 properties - special beaker, Special beakers – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1378

Advertising
background image

6.11 Subwindow Rack data

■■■■■■■■■■■■■■■■■■■■■■

1362

■■■■■■■■

tiamo 2.2 (ProcessLab)

Range

0 ... 235 mm

Rinse position

Rinse position for Lift 2. This lift position is used for rinsing the electrodes.

Range

0 ... 235 mm

Shift position

Shift position for Lift 2. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then the shifting will take place
at the current lift position. This means that the shift position must be selected
so that a safe movement across the entire rack is possible at any time.

Range

0 ... 235 mm

Special position

Special position for Lift 2. This additional definable position can be used e.g.
during pipetting so that the tip is just immersed in the sample solution.

Range

0 ... 235 mm

6.11.3.4

Properties - Special beaker

Tab: Configuration

Rack data [Edit] Properties... Rack data Special beakers

The table shows all the data of all the special beakers of the attached rack
in tabular form. Each special beaker can be assigned to any position on the
rack. The table cannot be edited directly.

Note

Special beakers should preferably be set at high rack positions so that
sample series can start from rack position 1. Rack positions which are
defined as special beakers are no longer available as sample positions. A
specific work position can be defined for each special beaker at Tower 1
and Tower 2 (if present). The shift, rinse and special positions of the
affected tower are taken from the general rack positions.

The following columns are shown in the table:

Special beaker

Number of the special beaker for the selected rack.

Rack position

Number of the rack position for the special beaker.

Advertising