5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

498

■■■■■■■■

tiamo 2.2 (ProcessLab)

Sensor

Selection of a sensor of type Metal electrode from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

I(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.

Titrando, 855
Range

-125.0 ... 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Range

-127 ... 127 µA (Increment: 1)

Default value

5 µA

Electrode test

on | off (Default value: off)

If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.