5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

965

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically when
the command has finished.

5.6.4.10.3

MEAS Cond - Measuring parameters

Tab: Method

MEAS Cond Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement

Measurement frequency

712
Selection

auto | 300 Hz | 2.4 kHz

Default value

auto

auto
The more suitable of the two frequencies is selected.

Type of measurement

The two measuring modes Measurement with drift control or Meas-
urement without drift control
can be selected.

Selection

Measurement with drift control | Measurement
without drift control

Default value

Measurement with drift control

Advertising