4 meas conc - additional measured values, Additional measured values – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 975

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

959

Measurement without drift control
If the option Measurement without drift control has been selected, then
the following parameters will be displayed:

Measuring time

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list. Max.
number of entries?

Range

0.1 ... 999999.0 s

Default value

2.0 s

Stop measured value

Stops when the preset measured value has been reached since the start of
the measurement.

Range

0 ... 1E+99

Selection

off

Default value

off

off
No stop.

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously.

Range

-20.0 ... 150°C

Default value

25.0

5.6.4.9.4

MEAS Conc - Additional measured values

Tab: Method

Commands MEAS Conc Properties... Additional measured

values

Command name

Name of the command.

Input

25 characters