Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 870

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

854

■■■■■■■■

tiamo 2.2 (ProcessLab)

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this check box is activated, then the measured value will be monitored and
any exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is triggered.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

-2000 mV

Lower hysteresis

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value lower
limit ok
is triggered.

Titrando, 855
Range

0.0 ... 2000.0 mV

Default value

0.2 mV

Upper limit

Upper limit of the measured value. If the measured value exceeds this limit,
then the event Measured value upper limit exceeded is triggered.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

2000.0 mV

Titrino
Range

-2000 ... 2000

Default value

2000

Upper hysteresis

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.