8 Manual Control

tiamo 2.2 (ProcessLab)

■■■■■■■■

1525

Range

1 ... 999 min

Selection

off

Default value

off

[Start]

Start heater. In the status display the current temperature is shown.

[Stop]

Stop heater.

Gas flow

Inlet

Select the inlet of the carrier gas.

Selection

Pump | Valve

Default value

Pump

Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.

Flow rate

874
Range

10 ... 150 mL/min

Default value

50 mL/min

Gas type

Gas type of the carrier gas.

Selection

Air | Nitrogen | Other gas

Default value

Air

Gas flow factor

Factor for the correct determination of the gas flow. This parameter is only
displayed for Gas type = Other gas. The values for frequently used carrier
gases are listed in the manual of the 774 and 874 Oven Sample Processor.

Range

1 ... 9.999

Default value

1.000

[Start]

Switch on gas flow. In the status display the running time since start is
shown.

[Stop]

Switch off gas flow.