Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 882

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

866

■■■■■■■■

tiamo 2.2 (ProcessLab)

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead.

If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | 859 Titrotherm

Default value

859 Titrotherm

Dosing device

Dosing device

Selection of the number of the dosing device (dosing unit) with which the
solution is to be dosed. All the dosing device connections which are possible
with the selected device type are displayed.

Selection

1 | 2 | 3 | 4

Default value

1

Solution

Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent dosing units are used, then a check will be
made in the method run to see whether the correct solution has been
attached to the selected dosing device. With non-intelligent dosing units,
only the cylinder volume is checked. At the start of the command, a check
is made of the working life, the validity of the titer and the GLP test interval
for the selected solution.

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input the sensor is connected to.

Selection

1 | 2

Default value

1

Sensor

Selection of a sensor of type Thermoprobe from the sensors configured in
the sensor table. The calibration data for the sensor will be adopted for the
determination.