5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

766

■■■■■■■■

tiamo 2.2 (ProcessLab)

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical, fast titrations.
User
Editing the individual titration parameters which affect the titration rate.

Control

Note

Will be displayed only if the User option is selected in the Titration
rate
list box.

Dynamics

The control range defines the measured value range before the given end-
point. It has a decisive influence on the titration rate and thus also on the
accuracy. In the control range, dosing is controlled by the Minimum rate.
The closer the endpoint, the slower the dosing until the Minimum rate has
been reached. The larger the control range, the slower the titration. Outside
the range the dosing takes place at the Maximum rate.

Titrando
Range

0.01 ... 200.0 µA

Default value

40.0 (slow), 10.0 (optimal), 5.0 (fast) µA

Titrino
Range

0.1 ... 200.0 µA

Default value

40.0 (slow), 10.0 (optimal), 5.0 (fast) µA