Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1552
8.4 Manual Control - Functions
■■■■■■■■■■■■■■■■■■■■■■
1536
■■■■■■■■
tiamo 2.2 (ProcessLab)
For Type = Stirrer:
■
0: Duration of the switch-on pulse fixed at 500 ms.
■
>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in 10
ms
Example:
Parameter 1 = 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on 225
ms, switched off 75 ms
Range
-214783647 ... 214783647
Default value
0
Start delay
Waiting time in seconds or minutes before the initiated action is started.
Range
0 ... 999999
Default value
0
Selection
s | min
Default value
s
Working time
Time in seconds or minutes for which the digital output is set to ACTIVE.
After the working time has elapsed, it is set to INACTIVE again.
Range
0 ... 999999
Default value
0.01
Selection
s | min
Default value
s
End delay
Waiting time after the working time has expired
Range
0 ... 999999
Default value
0
Selection
s | min
Default value
s
Status
Display of the current status.
Selection
ACTIVE | INACTIVE | ERROR
Time to finish
Display of the time to finish of the timer control.