6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1295

Note

This parameter is not saved on the data chip of intelligent exchange/dos-
ing units, i.e. this parameter will be empty if the data are written into the
solutions table from the IEU/IDU.

User

Short name of the user logged in during titer determination or when the
titer was entered manually. If work is not carried out using login then the
user logged in under Windows will be entered automatically.

Note

This parameter is not saved on the data chip of intelligent exchange/dos-
ing units, i.e. this parameter will be empty if the data are written into the
solutions table from the IEU/IDU.

History graph
The graph shows the last 10 titer values for the selected solution. The titer
values are shown in different colors in the history table:

Blue, if the titer has been determined automatically by a method.

Black, if the titer has been entered manually.

If limits have been defined then the warning limits will be shown in
orange and the intervention limits in red.

Note

The history graph can be copied into the clipboard by using the context-
sensitive menu item Copy.

[Limits]

Opens the dialog window Limits for titer, in which the warning and inter-
vention limits can be defined for the titer (see Chapter 6.6.3.5, page
1296)
. These limits apply only to the graph, no monitoring is carried out
during the titer determination.

[Delete history]

Deletes titer history.