5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

868

■■■■■■■■

tiamo 2.2 (ProcessLab)

Range

0.00000 ... 50.0000 mL

Default value

0 mL

Dosing rate

Rate (volume/time unit) at which the start volume is to be added. The max-
imum dosing rate depends on the cylinder volume of the dosing unit used.

Range

0.100 ... 166.000 mL/min

Selection

maximum

Default value

maximum

Pause

Pause

Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the dosing of a start volume. The pause follows at the end
of all the start conditions.

Range

0 ... 9999 s

Default value

0 s

5.6.3.9.4

TET - Titration parameters

Tab: Method

TET Properties... Titration parameter

Command name

Name of the command.

Input

25 characters

The parameters for the running of the titration are defined on this tab.

Dosing

Dosing rate

Rate (volume/time unit) at which dosing is carried out. The maximum dosing
rate depends on the cylinder volume of the dosing unit used.

Range

0.100 ... 166,000 mL/min

Default value

2.000 mL/min

Selection

maximum

Filter

Filter factor

Factor which determines the digital filtering of the measuring curve. The
higher this value, the more the noise is damped. The filtering has an influence
on the shape of the peak and the EP shift.