5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

914

■■■■■■■■

tiamo 2.2 (ProcessLab)

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855, 867
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos,
855 and 867.

5.6.4.5.3

MEAS Ipol - Measuring parameters

Tab: Method

MEAS Ipol Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control
can be selected.

Selection

Measurement with drift control | Measurement
without drift control

Default value

Measurement with drift control

Measurement with drift control
The measurement is carried out drift-controlled if this option is enabled.
The measurement is stopped as soon as the defined Signal drift or the
Stop measured value is reached or a waiting time has elapsed. The
corresponding parameters are not visible if this option is disabled.
Measurement without drift control
The measurement is carried out without drift control if this option is ena-
bled. The measurement continues until one of the stop criterions
Measuring time or Stop measured value is met. The corresponding
parameters are not visible if this option is disabled.