3 meas t - measuring parameters, Measuring parameters – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 955

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

939

Titrino
Selection

1 | 2 | diff.

Default value

1

Sensor

Selection of a sensor of type Temperature sensor or pH Electrode from
the sensors configured in the sensor table.

Selection

Sensor name | Temperature sensor | pH Elec-
trode

Default value

Temperature sensor

Stirrer

Stirrer

Selection of the stirrer.

Titrando, 855, 856, 867
Selection

1 | 2 | 3 | 4 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855, 856, 867
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos,
855, 856 and 867.

5.6.4.7.3

MEAS T - Measuring parameters

Tab: Method

MEAS T Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.