6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1331

Range

-2.147E+9 ... 2.147E+9 mV

Default value

-30 mV

Upper limit

Upper intervention limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

30 mV

Warning limits for cell constant (only for conductivity sensors)
on | off (Default value: off)

If this option is enabled then the values for the cell constant will be shown
in orange if the limits are exceeded.

Lower limit

Lower warning limit for the cell constant.

Range

0.001 ... 500 /cm

Default value

2.0 /cm

Upper limit

Upper warning limit for the cell constant.

Range

0.001 ... 500 /cm

Default value

400 /cm

Intervention limits for cell constant (only for conductivity sen-
sors)
on | off (Default value: off)

If this option is enabled then the values for the cell constant will be shown
in red if the limits are exceeded.

Lower limit

Lower intervention limit for the cell constant.

Range

0.001 ... 500 /cm

Default value

0.001 /cm

Upper limit

Upper intervention limit for the cell constant.

Range

0.001 ... 500 /cm

Default value

500 /cm