5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

786

■■■■■■■■

tiamo 2.2 (ProcessLab)

Ipol

10 µA

20 µA

30 µA

EP

50 mV

100 mV

150 mV

Dynamics

70 mV

100 mV

120 mV

Max. rate

Maximum rate with which iodine is generated outside the Control range.

Range

1.5 ... 2240.0 µg/min

Default value

1000.0 (slow), maximum (optimal), maximum
(fast) µg/min

Selection

maximum

Min. rate

This parameter defines the minimum rate with which iodine is generated
inside the Control range. The smaller the rate the slower the titration and
the higher the accuracy.

756, 831
Range

0.3 ... 999.9 µg/min

Default value

0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min

Selection

minimum

851, 852
Range

0.3 ... 999.0 µg/min

Default value

0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min

Selection

minimum

Stop criterion

Stop criterion

Switches off titration when the endpoint has been reached and the stop
criterion has been met.

756, 831
Selection

drift | rel. drift

Default value

rel. drift

851, 852
Selection

drift | rel. drift | time | off

Default value

rel. drift

Stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.