Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1141

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1125

Target

Tower

Selection of the tower of the Sample Processor for moving to the required
lift position. Both Tower 1 and Tower 2 can always be selected even if the
Sample Processor has only one tower.

Selection

1 | 2

Default value

1

Lift position

Entering the desired lift position in mm or selecting a lift position predefined
for the rack Work position, Shift position, Rinse position or Special
position
. A lift height of 0 mm corresponds to the Home position. Here
the lift is moved to the uppermost stop.

730, 778, 789, 814, 815, 855, 864
Range

0 ... 235 mm

Selection

Work position | Shift position | Rinse position |
Special position | Home position

Default value

Work position

774
Range

0 ... 100 mm

Selection

Work position | Shift position | Rinse position |
Special position | Home position

Default value

Work position

874
Range

0 ... 110 mm

Selection

Work position | Shift position | Rinse position |
Special position | Home position

Default value

Work position

Note

Please make sure that none of the lift positions exceed the maximum
stroke path given in the tower properties of the device.

Parameters

Lift rate

Selection of the speed for moving the lift.

Advertising