Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1372

Advertising
background image

6.10 Subwindow Sample solutions (TC conductivity)

■■■■■■■■■■■■■■■■■■■■■■

1356

■■■■■■■■

tiamo 2.2 (ProcessLab)

Selection

Date selection

Message
The message options can only be edited if monitoring is switched on.

Message by E-mail

on | off (Default value: off)

The message is additionally sent to the E-mail address defined under [E-
mail]
if this option is activated. The message is sent in text format.

[E-mail]

Pressing [E-mail] opens the Send E-mail (see Chapter 2.6.1, page 86)
window.

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted additionally to the
message defined above

Action
The actions can only be edited if monitoring is switched on.

If during monitoring it is found that the validity period has expired then one
of the following actions will be triggered automatically at Start test.

Selection

Record message | Display message | Stop deter-
mination

Default value

Display message

Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can choose whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the deter-
mination.
Stop determination
The running determination will be automatically canceled. The following
message must be confirmed with [OK].

Advertising