5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

594

■■■■■■■■

tiamo 2.2 (ProcessLab)

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can be
taken over and saved from existing measuring commands (i.e. MEAS) which
can also be displayed in curves under the designation Extern 1…3 and used
in formulas as variables 'Command name.EX1…3'-

[New]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 880)
.

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.3.3.9

MET U - Measured value window

Dialog window: Method

MET U Properties... Potentiometric evalua-

tion

Evaluation with measured value window [New]/[Properties] Measured

value window#

Measured value windows are regions on the measured value axis for which
different parameters for potentiometric evaluation can be defined. Only
those endpoints which lie within these windows and additionally meet the
parameters defined for each window will be recognized. Only one endpoint
will be recognized per window.

Lower limit

Lower limit for the measured value window.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

-2000 mV

Upper limit

Upper limit for the measured value window.