5 det ph - stop conditions, Stop conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 469

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

453

Titrando, 855
Range

0.1 ... 9999.9 µL

Selection

off

Default value

off

Dosing rate

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

5.6.3.2.2.5

DET pH - Stop conditions

Tab: Method

DET pH Properties... Stop conditions

Command name

Name of the command.

Input

25 characters