5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1040

■■■■■■■■

tiamo 2.2 (ProcessLab)

Range

-20.0 ... 150.0 °C

Default value

25.0 °C

5.6.5.6

CAL LOOP Conc

5.6.5.6.1

CAL LOOP Conc - Overview

Dialog window: Method

CAL LOOP Conc Properties... CAL LOOP Conc - 'Com-

mand name'

Loop command for the calibration of ion-selective electrodes, which is
used together with the command CAL MEAS Conc.

Appearance
The command has the following appearance:

Parameters
The parameters for the command CAL LOOP Conc are configured in the
following dialog window:

CAL LOOP Conc - Properties

Principle
For calibrating ion-selective electrodes (ISE), a linear correlation of the meas-
ured potential U

i

and the logarithm of the sought ion concentration c

i

cannot

always be assumed. In ion measurements, interfering ions have an influence
on the measurement and that is why this effect is perceptible by a flattening
of the curve, especially in low concentration ranges.