5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

451

Measured value acceptance

Note

Will be displayed only if the User option is selected in the Titration
rate
list box.

Measured value drift

Drift for the measured value acceptance during the titration.

Titrando, 855
Range

0.1 ... 999.0 mV/min

Default value

20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Titrino
Range

0.5 ... 999.0 mV/min

Default value

20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.

Titrando, 855
Range

0 (slow, optimal, fast) ... 999999 s

Default value

0 (slow, optimal, fast) s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed. If the waiting time has not been newly entered then a waiting time
that is suitable for the drift will be calculated automatically according to the
following equation: