Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 520

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

504

■■■■■■■■

tiamo 2.2 (ProcessLab)

Titrino
Range

0.5 ... 999.0 mV/min

Default value

20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.

Titrando, 855
Range

0 (slow, optimal, fast) ... 999999 s

Default value

0 (slow, optimal, fast) s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed. If the waiting time has not been newly entered then a waiting time
that is suitable for the drift will be calculated automatically according to the
following equation:

Titrando, 855
Range

0.1 ... 999999 s

Default value

38.0 (slow), 26.0 (optimal), 21.0 (fast) s

Titrino
Range

0.1 ... 9999 s

Default value

38.0 (slow), 26.0 (optimal), 21.0 (fast) s

Dosing of increments

Note

The following parameters will be displayed only if the User option is
selected in the Titration rate list box.