7 rack, 1 rack - overview, Rack – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1148

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1132

■■■■■■■■

tiamo 2.2 (ProcessLab)

Period of operation
The stirrer can be switched on for a particular period.

Range

0 ... 9999.9

Default value

10.0

Selection

s | min

Default value

s

5.6.7.7

RACK

5.6.7.7.1

RACK - Overview

Dialog window: Method

RACK Properties... RACK - 'Command name'

Command for initialization of the attached sample rack. The rack, lift
and robotic arm (if present) are reset, the rack code is read off and the
corresponding rack data is transferred to the Sample Processor.

Devices
This command can be executed with the following devices:

Sample Processor: 730, 774, 778, 789, 814, 815, 855, 864, 874

Appearance
The command has the following appearance:

Parameters
The parameters for the command RACK are configured in the following
dialog window:

RACK - Properties

Command variables
The following command variables are generated in the method run by the
command RACK and can be used in formulas under the designation 'Com-
mand name.Variable designation'
:

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

Advertising