5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

747

Range

0.1 ... 9.9 µL/min

Default value

minimum (slow), minimum (optimal), 5.0
(fast) µL/min

Stop criterion

Stop criterion

The titration us stopped after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be stopped
when a particular drift has been achieved or after a preset time.

Titrino
Selection

drift | time

Default value

drift

Titrando
Selection

drift | rel. drift | off | time

Default value

drift

drift
The titration is stopped when the stop drift has been reached.
rel. drift
The titration is stopped when the relative stop drift (drift value at time of
start + relative stop drift) has been reached.
off
The titration will not be stopped.
time
The endpoint must be exceeded for a certain time: the Delay time. The
same Delay time with different, very small volume increments (depending
on the volume of the exchange or dosing unit) means different switch-
off points.

Note

The Stop conditions will always stop the titration even if the stop criterion
has not been fulfilled or has been switched off.

Stop drift

Titration is stopped when the endpoint and the stop drift have been reached.
This parameter is only shown for Stop criterion = drift.

Range

1 ... 999 µL/min

Default value

20 µL/min