5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

894

■■■■■■■■

tiamo 2.2 (ProcessLab)

measured value can be evaluated as a break point for a small measured value
range.

Range

0 ... 1.0

Default value

0.3

Slope

Minimum difference between the slope before and after the break point.
The smaller the difference, the more break-points will be found.

Range

0.0 ... 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer break-points will be found.

Range

2 ... 20

Default value

5

Window

A range (window) can be defined on the measured value axis or on the time
axis. The break point evaluation will only be carried out in the defined win-
dow. Only the first break point in the defined window will be recognized.

Selection

Measured value | time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Time for the lower limit of the window.

Window = Measured value
Range

-20,000 ... -20,000 pH

Default value

20,000 pH

Window = Time
Range

0 ... 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Value for the upper limit of the window.

Window = Measured value
Range

-20,000 ... 20,000 pH

Default value

20,000 pH