Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 987

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

971

Range

0 ... 1.0

Default value

0.3

Slope

Minimum difference between the slope before and after the break point.
The smaller the difference, the more break-points will be found.

Range

0.0 ... 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer endpoints will be found.

Range

2 ... 20

Default value

5

Window

Selection

Measured value | time | off

Default value

off

off
A range (window) can be defined on the measured value axis or on the
time axis. The break point evaluation will only be carried out in the
defined window. Only the first break point in the defined window will be
recognized.

Lower limit

Value for the lower limit of the window.

Time for the lower limit of the window.

Window = Measured value
Range

0.0000 ... 2000.0 mS/cm

Default value

0.0000 mS/cm

Window = Time
Range

0 ... 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Value for the upper limit of the window.

Window = Measured value
Range

0.0000 ... 2000.0 mS/cm

Default value

2000.0 mS/cm

Advertising