5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

646

■■■■■■■■

tiamo 2.2 (ProcessLab)

5.6.3.3.5.9

MET Upol - Measured value window

Dialog window: Method

MET Upol Properties... Potentiometric evalua-

tion

Evaluation with measured value window [New]/[Properties] Measured

value window#

Measured value windows are regions on the measured value axis for which
different parameters for potentiometric evaluation can be defined. Only
those endpoints which lie within these windows and additionally meet the
parameters defined for each window will be recognized. Only one endpoint
will be recognized per window.

Lower limit

Lower limit for the measured value window.

Range

-200.0 ... 200.0 µA

Default value

-200.0 µA

Upper limit

Upper limit for the measured value window.

Range

-200.0 ... 200.0 µA

Default value

200.0 µA

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.

Range

0.1 ... 99.9 µA

Default value

2.0 µA

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.