6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1239

Comment on modification of sample data (live)

If this check box is activated then each time that a determination is modified
a modification reason and a modification comment must be entered; these
are saved in the determination and shown in the database in the subwindow
Information on the tab Sample. The reason and comments are also recor-
ded in the Audit Trail.

Selection

on | off

Default value

off
FDA default: on

6.2.2.5

Signatures

Tab: Configuration

Tools Security settings Signatures

On the tab Signatures the parameters for the electronic signatures can be
configured.

Settings according to 21 CFR 11

[Set]

Sets the parameters to default settings for FDA-compliant work.

Inactivity delay

If this check box is activated then the dialog window for signing will be
automatically closed when the entered time limit has expired.

Selection

on | off

Default value

off
FDA default: on

Range

1 ... 60 min

Default value

10 min
FDA default: 10 min

Remove password after signature

If this check box is activated then the password must be entered again after
each signature.

Selection

on | off

Default value

off
FDA default: on

Sign each determination separately

If this check box is activated then each determination contained in the
determination overview must be signed individually.

Selection

on | off

Default value

off
FDA default: on