5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

885

Command variables
The following command variables are generated in the method run by the
command MEAS pH and can be used in formulas under the designation
'Command name.Variable designation':

Identification

Description

.BP{x}.MEA

Measured value for the break point x (1...9) in the unit
of the measured value

.BP{x}.TEM

Temperature for the break point x (1…9) in °C

.BP{x}.TIM

Time for the break point x (1…9) in s

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.DBL

Total duration for the processing of the command in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.ENP

Electrode zero point of the sensor used for the com-
mand (dimensionless)

.ETE

End temperature (temperature following processing of
the command) in °C

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command
has never ended

.FP{x}.MEA

Measured value for the fixed endpoint x (1...9) in the
unit of the measured value

.FP{x}.TEM

Temperature for the fixed endpoint x (1…9) in °C

.FP{x}.TIM

Time in s until the fixed endpoint x (1...9) is reached

.IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

.ITE

Initial temperature (temperature before start conditions
are processed) in °C

.LP.CAx

Calculated value x (1…3) for the last measuring point
on the measuring point list

.LP.EXx

External value x (1…3) for the last measuring point on
the measuring point list