5 met u - stop conditions, Stop conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 601

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

585

tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

5.6.3.3.3.5

MET U - Stop conditions

Tab: Method

MET U Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.000 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop measured value

Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

Advertising