Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1092

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1076

■■■■■■■■

tiamo 2.2 (ProcessLab)

Dosing device

Dosing device

Selection of the dosing device to be used for the dosing. All the dosing device
connections which are possible with the selected device type are displayed.

Titrando

Selection

1 | 2 | 3 | 4

Default value

1

855

Selection

1 | 2 | 3

Default value

1

Titrino
Selection

internal D0

736, 751, 799
Selection

internal D0 | external D1 | external D2

Default value

internal D0

Solution

Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Tandem dosing

on | off (Default value: off)

If this option is enabled, then dosing will proceed continuously with two
dosing devices in combination. The second dosing device is dosing while the
first one is being filled and vice-versa. This parameter is displayed only for
Titrandos and 855.