3.6 Subwindow Run

■■■■■■■■■■■■■■■■■■■■■■

142

■■■■■■■■

tiamo 2.2 (ProcessLab)

The second field after Autostart shows the total number of determinations
per series that are still to be carried out (Setpoint counter). It contains in its
default settings the entry Sample table, which refers to the number of lines
of the working sample table. This field cannot be edited except when it is in
the status READY.

Note

The second field after Autostart can also be modified live for a series
that is currently running. The Live modifications window opens with
the context-sensitive menu item Modify autostart counter.

Range

1 ... 999

Default value

Sample table

Statistics

on | off (Default value: on)

This field is displayed only when a method has been loaded in the START
command of which the option Statistics has been enabled. The statistics
defined in the method can be enabled and disabled manually here (only in
the status READY).

The first field after Statistics shows the number of determinations already
carried out for the statistics (Actual counter). This field cannot be edited, but
when it is in the status READY and COND BUSY, it can be set to 0 with
the context sensitive menu item Delete statistics, whereby the already
calculated statistics data is deleted at the same time.

The second field after Statistics shows the number of determinations still
to be carried out for the statistics (Setpoint counter). In its default settings,
it contains the value Number of single determinations defined in the
START command. This field cannot be edited except in the status
READY and COND BUSY.

Note

When a new statistics series is started, the setpoint counter will be reset
to the value defined in the START command.