5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

784

■■■■■■■■

tiamo 2.2 (ProcessLab)

End point

EP at

Measured value for the endpoint.

756, 831
Range

-2000.0 ... 2000.0 mV

Default value

50.0 mV

851, 852
Range

-1250.0 ... 1250.0 mV

Default value

50.0 mV

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | user

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical, fast titrations.
user
Editing the individual titration parameters which affect the titration rate.

Control

Note

Will be displayed only if the User option is selected in the Titration
rate
list box.