Lower limit value exceeded.
upper
Upper limit value exceeded.
ok
Measured value is inside limits again.

[pH]/[U]

Measured value limit.

Meas.v. [pH]/Meas.val. [U]

Measured value at limit violation time.

Volume [mL]

Dosing volume at limit violation time.

Temp. [°C]

Temperature at limit violation time.

Action

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Stop determination
The running STAT command was quit, then the determination was
stopped.
Stop command (only Titrando)
The running STAT command was quit, then the next command was
started.
Wait for [Continue]
Reagent dosing in the current STAT command was interrupted until it
was continued by a manual command.
Wait for limit ok
Reagent dosing in the current STAT command was interrupted until the
monitored measured value was again within the limits (including hyste-
resis). Afterwards reagent dosing was resumed automatically.
none
No action was taken.