5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

1055

Selection

Input | Request | Template

Default value

Input

Input
If this option is selected, then the values for the conductivity standard
can be entered directly.
Request
If this option is selected, then first the current temperature of the solution
is measured after the start of the determination. Afterwards, the dialog
window Data Conductivity standard will appear (see Chapter
5.6.5.8.5, page 1056)
.
The method continues running after the data has
been entered.
Template
If this option is selected, then a conductivity standard can be selected,
and its data will be used to calculate the temperature coefficient for the
specified reference temperature.

Parameters for input
If the option Input has been selected, then the following parameters will
be displayed:

Reference temperature

Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.

Range

-20.0 ... 150.0°C

Default value

20.0

Conductivity

Input of the conductivity of the standard at the reference temperature.

Range

0.00000 ... 2000.00 mS/cm

Default value

11.66 mS/cm

Temperature coefficient

Temperature coefficient of the standard with the reference temperature.

Range

0.00 ... 9.99 %/°C

Default value

2.07 %/°C

Parameters for template
If the option Input has been selected, then the following parameters will
be displayed: