Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 1008

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

992

■■■■■■■■

tiamo 2.2 (ProcessLab)

Range

0.1 ... 1000.0 mAU/s

Default value

20.0 mAU/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated time for the maximum measured
value will be interpolated from the measuring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 1000.0 mAU/s

Default value

20.0 mAU/s

Break-point evaluation

Break-point evaluation

on | off (Default value: off)

If this option is enabled, then sharp, almost 90° changes in directions in the
measurement curve will be evaluated.

EP criterion

Measure of the minimum sharpness of the break-point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value alteration, even small changes in the
measured value can be evaluated as a break-point for a small measured value
range.

Range

0.0 ... 1.0

Default value

0.3

Slope

Minimum difference between the slope before and after the break-point.
The smaller the difference, the more break-points will be found.

Range

0.0 ... 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer endpoints will be found.

Range

2 ... 20

Default value

5