6 Configuration

tiamo 2.2 (ProcessLab)

■■■■■■■■

1303

Dosing rate Dosing port 2

Speed at which dosing is to be carried out at Dosing port 2. The maximum
dosing rate depends on the cylinder volume of the exchange unit used (see
below). When the function is carried out the speed will automatically be
reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Dosing rate Fill port

Speed at which dosing or filling is to be carried out at the Fill port. The
maximum dosing rate depends on the cylinder volume of the exchange unit
used (see below). When the function is carried out the speed will automat-
ically be reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Dosing rate Special port

Speed at which dosing is to be carried out at the special port. The maximum
dosing rate depends on the cylinder volume of the exchange unit used (see
below). When the function is carried out the speed will automatically be
reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

The maximum dosing rate depends on the cylinder volume of the
dosing unit used (see below):

Cylinder volume

Maximum rate

2 mL

6.67 mL/min

5 mL

16.67 mL/min

10 mL

33.3 mL/min

20 mL

66.6 mL/min

50 mL

166.0 mL/min

Note

Enter lower rates for high-viscosity liquids.