5 met ph - stop conditions, Stop conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 575

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

559

Titrino
Range

0.001 ... 9.999 mL

Default value

0.05 (slow), 0.10 (optimal), 0.20 (fast) mL

Dosing rate

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

5.6.3.3.2.5

MET pH - Stop conditions

Tab: Method

MET pH Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Advertising