10 kfc - fix endpoint evaluation, 7 brc, 1 brc - overview – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 809

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2 (ProcessLab)

■■■■■■■■

793

5.6.3.6.10

KFC - Fix endpoint evaluation

Dialog window: Method

KFC Properties... Additional evaluations Fix end-

point evaluation

[New]/[Property] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Water

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Water
Range

0.0 ... 999999.9 µg

5.6.3.7

BRC

5.6.3.7.1

BRC - Overview

Dialog window: Method

BRC Properties... BRC - 'Command name'

Command for coulometric titration for determining the bromine consump-
tion. Out of the amount of bromine consumed the bromine index (amount
of bromine consumed (in mg) per 100 g sample) and the bromine num-
ber
(amount of bromine (in g) per 100 g sample) can be calculated.

Devices
This command can be executed with the following devices.

Titrandos: 851, 852

Appearance
The command has the following appearance:

without conditioning

with conditioning